Mensor Hand-Held Thermometer CTH65I0
Are you buying in volume? Are you an OEM? Are you a reseller? Be sure to contact FLW directly for your special pricing.
Description
The all-purpose model CTH6500 hand-held thermometer, for superior mobile temperature measurement, is notable for its precision, flexibility and ease of handling. In addition to Pt100 resistance thermometers, it can also process signals from typical thermocouples. Thus temperatures from -200 ... +1,500 °C (-328 ... +2,732 °F) can be measured.
Through its high accuracy of 0.03 K in ranges from -100 ... +150 °C (-148 ... +302 °F), this instrument can also be used as a reference instrument in biotechnology, pharmaceutical and food industries. The CTH6500 is thus also ideal for all service and maintenance tasks.
Low-drift measuring amplifiers ensure small measurement errors, while easy-to-use adjustment features considerably simplify adjustments and calibrations:
Applications:
Through its high accuracy of 0.03 K in ranges from -100 ... +150 °C (-148 ... +302 °F), this instrument can also be used as a reference instrument in biotechnology, pharmaceutical and food industries. The CTH6500 is thus also ideal for all service and maintenance tasks.
Low-drift measuring amplifiers ensure small measurement errors, while easy-to-use adjustment features considerably simplify adjustments and calibrations:
- Calibration by code for fast setting of standard probes via identification numbers
- Physical calibration of probe and display at one, two or three different temperatures
Applications:
- Calibration of thermometers
- Temperature measurement for the purposes of quality assurance
- Measurements in service and maintenance applications
- Long-term monitoring and online documentation
- High accuracy of 0.03 K with Pt100
- One- and two-channel versions
- Connection possibilities for various probe types
- Intrinsically safe version Ex ib IIB T4 Gb
Product Specifications
General Specifications
Probe type: | Pt100 |
---|---|
Measuring input: | 1 or 2 |
Pt100 Measuring Range: | -200 ... +600 °C (-392 ... +1,112 °F) |
Resistance thermometer type Pt100 Accuracy: | 0.03 K for -50 ... +199.99 °C (-58 ... +394.98 °F) 0.05 K for -200 ... -50.01 °C (-328 ... -58.02 °F) otherwise 0.05 % of reading |
Measuring rate: | 4/s ("fast"); 1/s ("slow") |
Operating temperature: | 0 ... 40 °C (32 ... 104 °F) |
Storage temperature: | -10 ... +50 °C (14 ... 122 °F) |