Mensor Temperature Dry-Well Calibrator CTD9100-COOL
Are you buying in volume? Are you an OEM? Are you a reseller? Be sure to contact FLW directly for your special pricing.
Description
Nowadays, fast and simple testing of thermometers is a "must" when it comes to the operational reliability of machines and plants. The portable calibrators of the CTD9100 family are particularly suited to local calibration tasks. They are extremely userfriendly. Due to their compact design and their low weight, the instruments can be taken and used almost anywhere.
The CTD9100-COOL calibrators work with Peltier elements and can therefore reach test temperatures below ambient temperature. Due to the feature of active cooling, they are frequently used in the bio-pharmaceutical and food industries.
Applications:
The CTD9100-COOL calibrators work with Peltier elements and can therefore reach test temperatures below ambient temperature. Due to the feature of active cooling, they are frequently used in the bio-pharmaceutical and food industries.
Applications:
- Easy on-site calibration
- Power generation
- Measurement and control laboratories
- Machine building
- Various temperature ranges
- Measurement uncertainties from 0.15 ... 0.8 K
- Compact design
- Simple operation
Product Specifications
General Specifications
Model CTD9100-COOL | Model CTD9100-165 | |
Temperature range: | -55 ... +200 °C (-67 ... +392 °F) | -35 ... +165 °C (-31 ... +329 °F) |
Accuracy: | 0.15 ... 0.3 K | 0.15 ... 0.25 K |
Stability: | ±0.05 K | ±0.05 K |
Resolution: | 0.01 up to 100 °C, then 0.1 (0.01 up to 212 °F, then 0.1) | 0.01 up to 100 °C, then 0.1 (0.01 up to 212 °F, then 0.1) |
Axial Homogeneity Temperature Distribution: | < 0.04 K at 200 °C (392 °F) | < 0.04 K at 100 °C (212 °F) 0.06 K up to 165 °C (329 °F) |
Radial Homogeneity Temperature Distribution: | dependent on temperature, temperature probes and their quantity | dependent on temperature, temperature probes and their quantity |
Heating time: | approx. 10 min from 20 to 200 °C (from 68 °F to 392 °F) | approx. 25 min from 20 to 165 °C (X approx. 35 min) (from 68 °F to 329 °F) |
Cooling time: | approx. 10 min from +20 to -20 °C (from 68 °F to -4 °F) | approx. 15 min from +20 to -20 °C (X approx. 35 min)(from 68 °F to -4 °F) |
Stabilisation time: | dependent on temperature and temperature probe | dependent on temperature and temperature probe |