• Teledyne Hastings DAVC-6 Digital Active Vacuum Controller, 0.133 to 133 Pa, DAVC-6-01-05

Teledyne Hastings DAVC-6 Digital Active Vacuum Controller, 0.133 to 133 Pa, DAVC-6-01-05

$385.00
Are you buying in volume? Are you an OEM? Are you a reseller? Be sure to contact FLW directly for your special pricing.  
Quantity:
Availability:
PART NUMBER:
DAVC-6-01-05
Shipping Weight:
1.00 LBS
Shipping:

FREE Ground Shipping

Description

The Teledyne Hastings Instruments' DAVC (Digital AVC) is a microprocessorbased vacuum gauge. In addition to the linear analog outputs (0-1 VDC, 0-5 VDC, 0-10 VDC, 4-20 mA, or 0-20 mA), pressure measurements can be accessed via the RS232 digital output. Digital readings can be provided in either Torr, mBar, or Pascal units.

A precision A/D converter, along with the microprocessor, measures the thermocouple vacuum gauge tube's output and converts the measurement to a pressure reading employing the gauge tube's well-defined output/pressure function.

The DAVC is a continuous-reading instrument for vacuum measurement and process control applications. The instrument is packaged in a small aluminum case with controls and electrical "D" connector located on the top.

DAVC - Digital Vacuum Controller:
The DAVC includes several enhanced features which makes it an attractive solution for many vacuum control automation applications. The DAVC can provide linear analog output (voltage or current) for integration of vacuum measurement with a PLC or other data acquisition system. In addition, the DAVC provides RS232 communication for direct connection of the instrument to a computer.

Features:
  • Linear Analog Output:
    • 0 - 1 VDC / 0 - 5 VDC / 0 - 10 VDC
    • 0 - 20 mA / 4 - 20 mA
  • Nonlinear Amplified Tube Output:
    • 0 ? 1 VDC
  • Digital Output ? RS232
  • Operates Hastings Most Popular Gauge Tubes
  • Compatible with Calibration Reference Tubes
  • 11 ? 32 VDC
  • TTL Setpoint for Process Control (w/LED)
  • CE Compliant: EN61326
Applications:
  • Vacuum Furnaces
  • Plasma Coating
  • Refrigeration / HVAC
  • Power/Transformer Fill & Maintenance
  • Cryogenics/Vacuum Insulation
  • Freeze Drying
  • Accelerators
Benefits:
  • Proven Reliability / Less Downtime
  • Easy Installation and Operation
  • Low Cost
  • Rugged
  • Small & Lightweight

Product Specifications

General Specifications

Range:0.133-133 Pa
Operating Temperature:-20-70° C
Accuracy:± (15% of reading + 1 mTorr) (max)
Input Power:11 - 32 VDC, 0.7 W
I/O Connector:15-pin High-density "D", Male
Gauge Tube Connector:8-pin Octal
Cable (CB-DAVC-1.5m):1.5 m I/O Cable
(Power, Linear Output, Digital I/O, Set point)
Serial Communications:RS232, 9600 - 19200 baud
Set Point Output:Open-collector Transistor (80 mA @ 30 VDC)
Pressure Units:Torr
Linear Output:4 - 20 mA Linear

Literature & Docs