• DH-Budenberg High Pressure Deadweight Tester CPB5000HP

DH-Budenberg High Pressure Deadweight Tester CPB5000HP

Contact FLW for price and availability.

USD
Are you buying in volume? Are you an OEM? Are you a reseller? Be sure to contact FLW directly for your special pricing.  
Quantity:
Availability:
PART NUMBER:
CPB5000HP
Shipping Weight:
75.00 LBS

Description

Applications:
  • Primary standard for defining the pressure scale in a range up to 5,000 bar
  • Reference instrument for factory and calibration laboratories for the testing, adjustment and calibration of pressure measuring instruments
  • Complete, stand-alone system, also suitable for on-site use
Special Features:
  • Total measurement uncertainty to 0.02% of measured value
  • Dual-range piston-cylinder system 2,600 bar with fully automated changing between ranges
  • Factory calibration includes traceability to national standards, as standard; with DKD/DAkkS calibration possible as an option
  • High long-term stability with recommended recalibration cycle every five years
  • Masses manufactured from stainless steel and aluminum, can be adjusted to local gravity


Proven Primary Standard

Pressure balances are the most accurate instruments available on the market for the calibration of electronic or mechanical pressure measuring instruments. The direct measurement of the pressure (p= F/A), as well as the use of high-quality materials enable a very small measurement uncertainty, in conjunction with an excellent long-term stability of fiver years. The pressure balance has therefore been used for years in factory and calibration laboratories in industry, national institutes and research laboratories.

Stand-alone Operation

Due to its integrated pressure generation and the pure mechanical measuring principle, the model CPB5000 is ideal for on-site use for maintenance and service.

Basic Principle

Pressure is defined as the quotient of force and area. The core component of the CPB5000 is therefore a very precisely-manufactured piston-cylinder system, which is loaded with masses in order to generate the individual test points. The masses applied are proportional to the target pressure and this is achieved through optimally graduated masses. As standard, these masses are manufactured to the standard gravity (9.80665 m/s²), though they can be adjusted to a specific location and also DKD/DAkkS calibrated.

Easy Operation

The setting of the pressure is made via an integrated pump. For fine adjustment, a very precisely-controllable spindle pump is fitted, with a spindle running within it. As soon as the measuring system reaches equilibrium, there is a balance of forces between the pressure and the mass load applied. The excellent quality of the system ensures that this pressure remains stable over several minutes, so that the pressure value for comparative measurements can be read without any problems, or also so that more complex adjustments can be carried out on the item under test.

Product Specifications

Pressure Ranges:up to 70,000 psi (5000 bar)
Accuracy:0.020% of reading

Literature & Docs

Videos

How Deadweight Testers Work (05:29)
Shop deadweight testers at http://store.flw.com/categories/deadweight-testers.html or call (800) 576-6308. How a deadweight tester works to generate pressure for calibration and demonstration of the Mensor Deadweight tester model CPC5000 and the CPU5000 Control unit that automatically senses environmental conditions and adjusts for them.
  • How Deadweight...
    Shop deadweight testers at http://store.flw.com/categories/dea...